
GAS ANALYZER 359
5. O2S Cal Warning
The Oxygen Calibration Warning message indicates that the oxygen sensor did not
measure the minimum voltage required for a fixed oxygen setting. This occurs for the
following reasons: the oxygen sensor connector is not connected, not enough air flow is
reaching the oxygen sensor, or the oxygen sensor has reached the end of its useful life. If
the first two conditions are met, then replace the oxygen sensor.
6. NOx Zero Warning
The NOx Zero Warning message indicates that the sensor hardware is not working
properly. Re-run the NOx Zero procedure under the <MAINTENANCE> menu. If the
message returns, then an internal problem exists.
7. NOx Cal Warning
The NOx Calibration Warning message indicates that the NOx sensor did not measure the
minimum voltage required for a minimum NOx setting. This occurs for the following
reasons: the NOx sensor is connected improperly, the wrong bottle value for NOx was
entered during calibration, or the NOx sensor has reached the end of its useful life. If the
first two conditions are met, then replace the NOx sensor. If the bottle doesn't contain any
NOx, the NOx bottle value must be set to 0. (See Gas Analyzer Calibration on page 352).
8. Single Pt Cal Warn
The Single-point Calibration Warning occurs under two different operations. If this
message occurs during a zeroing operation, then the Oxygen Sensor is suspect. Check the
oxygen sensor connector. Re-zero, pressing 5 under the <GAS ANALYZER> menu.
This may need to be repeated. If the Single-point Calibration Warning occurs during a
Single-point Calibration, it indicates that one or more of the following gases (HC, CO,
CO2, or NOx) calibrated with more than 5% difference from the last Single-point
Calibration. Check the bottle values entered and perform a Single-point Calibration again.
If the bottle doesn't contain any NOx, the NOx bottle value must be set to 0. (See Gas
Analyzer Calibration on page 352).
9. Two Pt Cal Warn
The Two-point Calibration Warning occurs when one or more of the following gases (HC,
CO, or CO2) calibrated more than 5% difference from the last calibration. Recalibrate the
unit, checking the bottle values for correct concentrations. If the warning continues to
occur, refer to the Introduction section on Warranty and Repair information.
10. Low Flow
The Low Flow message indicates an obstruction in the sample system. Remove sample
hose from the analyzer. If the Low Flow condition goes away, check the hose for kinks or
obstructions. Blow air through it to remove possible obstructions. If the Low Flow
condition still exists, check the filters. If the Low Flow condition exists with the filter
bowl separated, then the unit has an internal problem. If the filter bowl is full of water,
stop testing and remove power immediately, inspect the check valve and drain filter, and
assure that the drain pump is operational.
Kommentare zu diesen Handbüchern